Display/Error Message
1. Error 1 / Er 1
Situation: Ambient temperature is out of operating rang 10-40 degree celcius
Solution: Allow the thermometer to remain in room temperature (10-40degree celcius) for 30 minutes
2. Error 2 / Er 2
Situation: Sensor Error - Sensor function of the thermometer no longer working well.
Solution: Shake the thermometer to check if there are any loose part or change the batteries
If it still does not function well, check the battery shell on the warranty. Contact us feedback@littlelives.com and provide the serial number and date of purchased.
3. Error 3 / Er 3
Situation: Memory Error
Solution:
You may try these steps to
1. Press POWER and the display will show the last records
2. Press MEM key to switch records shown,
"LCD displays test records and record number
3. Press the MODE key to switch mode
If it still does not function well, check the battery shell on the warranty. Contact us feedback@littlelives.com and provide the serial number and date of purchased.
4. Error 4 / Er 4
Situation: The temperature of the sensor body is unstable.
Solution: Place the thermometer in the measurement room to be stable for 30 minutes.
5. High / Hi
Situation: Temperature measured is higher than normal human temperature range
(Body above 42.2 degree celcius: Surface above 60 degree celcius.
Solution:
Check that the environment, should not be in a hot area.
Student who just came back from hot sun or was under hot area, allow student to cool down
Retake the temperature again once student has cool off or the area is not in a high temp environment